Differential detection of dual traps improves the spatial resolution of optical tweezers.
Proc Natl Acad Sci U S A 103(24):9006-11 (2006) PMID 16751267 PMCID 1482556
The drive toward more sensitive single-molecule manipulation techniques has led to the recent development of optical tweezers capable of resolving the motions of biological systems at the subnanometer level, approaching the fundamental limit set by Brownian fluctuations. One successful approach has been the dual-trap optical tweezers, in which the system of study is held at both ends by microspheres in two separate optical traps. We present here a theoretical description of the Brownian limit on the spatial resolution of such systems and verify these predictions by direct measurement in a Brownian noise-limited dual-trap optical tweezers. We find that by detecting the positions of both trapped microspheres, correlations in their motions can be exploited to maximize the resolving power of the instrument. Remarkably, we show that the spatial resolution of dual optical traps with dual-trap detection is always superior to that of more traditional, single-trap designs, despite the added Brownian noise of the second trapped microsphere.
DOI: 10.1073/pnas.0603342103
Version: za2963e q8zaa q8zb9 q8zc2 q8zdb q8zec q8zf3 q8zg2