In Vivo Wear Rate of the Birmingham Hip Resurfacing Arthroplasty
J Arthroplasty 24(6):6 (2009) PMID 18757168
The wear of 8 femoral and 2 acetabular components of the Birmingham Hip Resurfacing (Midland Medical Technologies Ltd, Birmingham, UK) was measured using a coordinate measuring machine (CMM5; SIP, Geneva, Switzerland). After a median in situ time of 13 months, measurements of the Birmingham Hip Resurfacing femoral heads revealed a median volumetric wear of 2.9 mm^3. In one case with a cup abduction angle of 70^o, a significantly higher wear of 17.8 mm^3 was found. The 2 acetabular components showed volumetric wear of 5.4 and 27.6 mm^3 after 14 and 15 months in vivo. The wear rates of the investigated femoral components were somewhat, but not significantly, higher than the previously investigated 28-mm Metasul heads (median, 0.8 mm^3; Zimmer GmbH, Winterthur, Switzerland).