Advanced search×

Arterial line for monitoring SpO2 in patients with ischemic peripheries.

J Clin Monit 24(4):279-81 (2010) PMID 20635123

Monitoring the oxygenation status of patients with poor peripheral perfusion or ischemic peripheries is challenging in view of unreliable or unrecordable pulse oximeter data. In this article we describe a very simple and innovative technique of using the arterial line for reliable recording of arterial oxygen saturation (SpO(2)) in such patients. We conclude that the arterial line can be used as an extension of the artery and SpO(2) may be reliably measured using the arterial line in such patients as long as the blood in the arterial tubing is pulsatile and a good contact is ensured between the arterial tubing and the sensor of the pulse oximeter.

DOI: 10.1007/s10877-010-9245-1
Version: za2963e q8za2 q8zb0 q8zc2 q8zd7 q8ze1 q8zfc q8zgd

Similar articles you may find interesting…

  1. Superior vena caval aneurysm

    Interact Cardiovasc Thorac Surg 2(3):3 (2003) PMID 17670060

  2. Emergency surgery after percutaneous transmitral commissurotomy: Operative versus echocardiographic findings, mechanisms of complications, a...

    J Thorac Cardiovasc Surg 130(3):5 (2005) PMID 16153927

    In 1388 cases of valvotomy, complications necessitating urgent surgery occurred in 31 cases (2.2%). Acute mitral regurgitation occurred in 23 cases (74.2%), and cardiac tamponade occurred in 8 cases (25.8%). Mitral regurgitation was due to leaflet tearing in all cases: anterior leaflet in 20 cases a...
  3. Low cost continuous femoral nerve block for relief of acute severe cancer related pain due to pathological fracture femur.

    Indian J Palliat Care 16(3):180-2 (2010) PMID 21218010

    We describe two such cases where severe pain was adequately relieved in the acute phase. Continuous femoral nerve block was used as an efficient, cheap and safe method of pain relief for two of our patients with pathological fracture femur. This method was proved to be quite efficient in decreasing...