A new kaolin-based haemostatic bandage compared with manual compression for bleeding control after percutaneous coronary procedures.
Eur Radiol 21(8):1687-91 (2011) PMID 21476127
Bleeding and vascular access site complications are an important cause of morbidity after percutaneous femoral procedures. New haemostatic dressings have been developed to control heavy bleeding. To evaluate the efficacy of a new kaolin-based haemostatic bandage for femoral artery closure after diagnostic or interventional procedures compared with manual compression. The first pilot European trial using this haemostatic bandage was performed at the in Milan, Italy. Two-hundred patients (71% male, mean age 66 ± 11 years) undergoing angiography or PCI via a femoral approach were randomised to the haemostatic bandage (n = 100) or manual compression (n = 100) following sheath removal. The mean active clotting time (ACT) at haemostasis was 146 ± 24 s (range 98-198 s). Haemostasis was achieved in 5.4 ± 1.5 min with the bandage vs 25 ± 15 min after manual compression, p
DOI: 10.1007/s00330-011-2117-3
Version: za2963e q8za0 q8zb6 q8zcf q8zd1 q8ze4 q8zf1 q8zg2